The new user-friendly touch screen-based interface makes it a self-explanatory unit with interactive learning on the go, including electrode placement, running EMG graph, EMG biofeedback Games and Templates as well as many other helpful tools.
- Pre-set programmes for Incontinence treatment (including a special programme for the treatment of incontinence by stimulating the nervous tibialis posterior with self-adhesive electrodes) and neuro-muscular Rehabilitation.
- Supplied with Rubber sleeve with stand and lanyard attachment option.
- Designed for Physiotherapist in mind to be used as a portable tool. Suitable as a personal trainer for in-clinic or home use (under guidance).
- Single channel EMG (Biofeedback) combined with 1 channel of NMES (stimulation).
- Single channel ETS (EMG triggered stimulation) with single channel stimulation.
- Simple single-phase or multiphase operation: multiphase allows EMG/ETS/STIM to be combined in one programme.
- Multilingual full colour display screen and voice prompts for Biofeedback.
- Manual and Automatic threshold control.
- EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
- Standalone device or used in conjunction with established multilingual software.
- Up to 10 meters of wireless (Bluetooth) connection for NeuroTrac Software.
- For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day-to-day basis providing average readings on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
- Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
- 1. EMG
- Single channel EMG
- EMG Range: 0.2 to 2000 μV RMS (continuous)
- Sensitivity: 0.l μV RMS
- Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
- Selectable Bandpass filter – 3db Bandwidth,
- a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% – Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% – Reading above 235 microvolts
- b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- Notch filter: 50 Hz (Canada 60Hz) – 33 dbs (0. 1% accuracy)
- Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- Work / Rest periods: 2-99 seconds
- Number of Trials: 2-99
- NMES, TENS and ETS
- Single channel Stimulator
- Amplitude: 0-90 mA into 500 Ohm load – actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 86 mA, at 1500 Ohms load the maximum will be limited to 65 mA.
- Type: Constant current, maximum output voltage 180 Volts +10 / -30 Volts
- Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
- Pulse width selection: 50-450 μS in predefined, 50-330μS in custom (10% accuracy)
- Pulse rate selection: 2-100 Hz (5% accuracy)
- Work / Rest periods: 2-99 seconds
- Time 1 – 99 minutes
- Ramp up time: 0.1 – 9.9 seconds
- Preset and user programmable treatment Programs
- Automatic output shut off with detection of open electrode above 10 mA (+/- 2mA).
Battery:
Battery set: 4 x 1.5V, AAA battery.
Low battery indication at 4.2-4.4 volts +/- 0.2 volts, automatic shut off when voltage drops below the low indication. Replace the batteries immediately!
The device goes off automatically when not in use (energy saving): for example when in settings and no key is pressed over 1 minute when in stimulation mode (home screen) and all channels are 0mA.
Expected average battery set life [of standard 850 mAh, alkaline]:
14-18 hours in STIM, 28 hours in EMG mode.
Environmental Conditions for use:
+5 to +40 degrees Centigrade. 15-90% Humidity.
Environmental conditions for storage & transport:
-25 to +70 degrees Centigrade. 15-90% Humidity.
Physical:
Dimensions: Length 150mm, Width 89mm, Depth 35mm
Weight: 140g (without batteries)